Technical Parameters
Design volume | 10/25/50/100/200/500 mL |
Maximum operating pressure | 49bar |
Maximum operating temperature | 500℃ |
Sealing form | Line-type self-seal |
Material of reactor | 316L Stainless steel |
Material of liner | None |
Precautions in Use of Hydrothermal Reactor
1、 Before using the hydrothermal reactor, please carefully check the saturated vapor pressure of the solvent at the reaction temperature, and confirm that the maximum allowable operating pressure of the hydrothermal reactor will not be exceeded.
2、 For the HTG/HTS series hydrothermal reactor, as the liner is made of PTFE, which is compressible under the high-temperature and high-pressure, the reactor lid and set screw should not be overtightened, so as to avoid severe deformation of the liner caused by thermal expansion under high temperature and high pressure.
3、 In case the operating temperature of the hydrothermal reactor exceeds 150 ℃, it is recommended to apply a layer of high-temperature lithium base grease on the thread before tightening the reactor lid to prevent the stainless steel thread from galling.
4、 It is prohibited to use the hydrothermal reactor under the following circumstances:
a. The solvent has a boiling point lower than 60 ℃ (e.g., ether, acetone, and dichloromethane)
b. Gas will be generated in the reaction
c. Materials with flammable, explosive or highly or extremely toxic properties
Factory and package